RecallDepth

Boehringer Replacement Gauge Model 9071, Lot Number 200405.

This recall has been terminated (originally issued July 20, 2004).

Company
Boehringer Laboratories, Inc.
Recall Initiated
May 19, 2004
Posted
July 20, 2004
Terminated
September 14, 2004
Recall Number
Z-1232-04
Quantity
9 units
Firm Location
Norristown, PA
Official Source
View on FDA website ↗

Reason for Recall

The suction regulator incorporates a gauge for the verification of suction levels. The particular lot of gauges used in the production of the regulators exhibits a potential to develop a leak and cause the delivered suction to be inaccurate.

Distribution

The recalled product was distributed to the following states: AL, FL, MA, MI, OR, PA, TN,

Lot / Code Info

Lot Number 200405

Root Cause

Other

Action Taken

All direct accounts were contacted by the recalling firm''s Customer Service Representative to determine the location of the affected product. The CSR arrange for replacement proudct.

More recalls by Boehringer Laboratories, Inc.

View all recalls by this company →