3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424.

Company
Daavlin Distributing Company
Recall Initiated
September 1, 2003
Posted
July 30, 2004
Terminated
July 24, 2007
Recall Number
Z-1233-04
Quantity
2 units distributed domestically.
Firm Location
Bryan, OH

Reason for Recall

Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.

Distribution

Nation wide plus Canada, the Philippines, Saudi Arabia, Turkey & Germany

Lot / Code Info

All Touch Control System, Model #PASNBSM-2424 units manufactured between 6/5/03 and 4/30/04.

Root Cause

Other

Action Taken

The firm has software patches to correct the programming problems as they are reported by the users. These patches are being installed either by a service representative for the manufacturer or by the customer