ViewForum Review Station Part Numbers: 9896 507 71830, 9896 507 71831, and 9896 507 71832.
This recall has been terminated (originally issued August 4, 2004).
- Recall Initiated
- June 8, 2004
- Posted
- August 4, 2004
- Terminated
- February 9, 2006
- Recall Number
- Z-1308-04
- Quantity
- 27
- Firm Location
- Bothell, WA
- Official Source
- View on FDA website ↗
Reason for Recall
When reviewing images in the module before sending them to PACS or printer the order can be erratic and out of sequence and may cause confusion with the interpretation of the examination.
Distribution
Philips Medical Systems North American Company distributes product to medical facilities nationwide.
Lot / Code Info
83616, 103380, 102578, 105315, 504608, 104095, 104900, 103620, 505119, 105295, 103540, 105735, 105608, 105493, 101585, 225694, 102492, 504979, 102101, 105290, 105486, 505784, 104079, 504925, 504952, 505815, 505960.
Root Cause
Other
Action Taken
The firm mailed letters via certified mail to their affected customers beginning on June 8, 2004. The letters instruct their customers to modify the default display on the ViewForum to exam/series date and time until a service representative can install the upgraded software R3.2.