ViewForum Review Station Part Numbers: 9896 507 71830, 9896 507 71831, and 9896 507 71832.

This recall has been terminated (originally issued August 4, 2004).

Company
Philips Medical Systems Sales & Service Region No. America
Recall Initiated
June 8, 2004
Posted
August 4, 2004
Terminated
February 9, 2006
Recall Number
Z-1308-04
Quantity
27
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

When reviewing images in the module before sending them to PACS or printer the order can be erratic and out of sequence and may cause confusion with the interpretation of the examination.

Distribution

Philips Medical Systems North American Company distributes product to medical facilities nationwide.

Lot / Code Info

83616, 103380, 102578, 105315, 504608, 104095, 104900, 103620, 505119, 105295, 103540, 105735, 105608, 105493, 101585, 225694, 102492, 504979, 102101, 105290, 105486, 505784, 104079, 504925, 504952, 505815, 505960.

Root Cause

Other

Action Taken

The firm mailed letters via certified mail to their affected customers beginning on June 8, 2004. The letters instruct their customers to modify the default display on the ViewForum to exam/series date and time until a service representative can install the upgraded software R3.2.