Product is a ChoICE 182 cm Guide Wire, intended to facilitate placement of balloon dilatation catheters and other therapeutic devices during Percutaneous Transluminal Coronary Angioplasty (PTCA) procedures. The ChoICE guide wire contains a one-piece stainless steel core, which extends throughout the length of the wire and provides steerability and torqueability. The wire is tapered and formed at the distal tip. A radiopaque platinum spring coil surrounds the distal tip of the core wire. The ChoICE Guide Wire is a single use, sterile (EO), device packaged in a carrier tube, which is held in a coil by plastic clips. The carrier tube assembly of the guide wire is sealed into a pouch. One side of the pouch is clear polyethylene/polyester laminate and the reverse side is Tyvek. The pouch assemblies are packaged in a five-pack product box. A flushing tool is included in the package to facilitate hydration of the coating with heparanized saline before use. 1, page 1, is a copy of the pouch label. Exhibit 1, page 2 is a copy of the five-pack product box label. Exhibit 2 is a copy of the Directions for Use.
This recall has been terminated (originally issued February 27, 2003).
- Company
- Symbiosis Corp.
- Recall Initiated
- November 19, 2002
- Posted
- February 27, 2003
- Terminated
- April 15, 2008
- Recall Number
- Z-0592-03
- Quantity
- 19/5-pack units
- Firm Location
- Miami, FL
- Official Source
- View on FDA website ↗
Reason for Recall
Product is mislabled as a ChoICE Floppy Guide Wire when it is actually a ChoICE ES(Extra Support) Guide Wire.
Distribution
Product was distributed to hospitals in FL, MI, GA, OR, ID AND CT. 05.
Lot / Code Info
Individual pouch codes are coded catalog # 12132-01, batch 4785750 expiration date 2004-05 and the five pack outer boxes are coded: Catalog No. 12132-01, batch 4806821 expiration date 2004-5.
Root Cause
Other
Action Taken
Recall Letters issued 11/19/2002 via Federal Express to 7 hospital accounts. Consignees were requested to fill out a response form and return product to their Quincy, MA facility.
More recalls by Symbiosis Corp.
Medi-Tech Magic Torque Glidex hydrophilic coated guidewire, .035 inch/180 cm and .035 inch/260 cm le...
Mar 20, 2003
Scimed PT Graphix Intermediate PCTA Guide Wire with ICE Hydrophilic Coating, 300 cm length.
Feb 27, 2003
Scimed ChoICE PT Extra Support PTCA Guide Wire with ICE Hydrophilic Coating, 300 cm length
Feb 27, 2003