7F NeuroFlo Catheter, Model 1027 The NeuroFlo Catheter is a multi-lumen device with two balloons mounted near the distal tip for the treatment of cerebral ischemia resulting from symptomatic vasospasm following aneurysmal subarachnoid hemorrhage. The NeuroFlo Catheter is under Humanitarian Device Exemption and not commercially released under a premarket notification or premarket approval.

Company
Zoll Circulation, Inc.
Recall Initiated
April 1, 2014
Posted
December 3, 2014
Terminated
November 26, 2014
Recall Number
Z-0477-2015
Quantity
1 device
Firm Location
San Jose, CA

Reason for Recall

After aquisition of assets of CoAxia, Zoll discovered that one lot of NeuroFlo Catheters was manufactured by another firm and an HDE supplement listing as the owner of the product was not filed.

Distribution

One location in NC

Lot / Code Info

Model 1027, Catalog number 8700-000775-01, lot 0021.

Root Cause

Other

Action Taken

Hospital was notified in April 2014 of the recall by UPS.