RecallDepth

Burstberry Laser Projection Series

Company
Gk Photonics Inc
Recall Initiated
July 10, 2019
Recall Number
Z-2097-2019
Firm Location
Glendale, CA

Reason for Recall

The User Remote Interlock Dongle provided with the E-Stop Remote is not sufficient to fully comply with the FDA's Manual Reset Mechanism requirement for laser projectors and could unintentionally expose the user and/or observers to high power laser radiation. It needs to be replaced for safety reasons.

Distribution

Unknown

Lot / Code Info

Burstberry

Root Cause

Radiation Control for Health and Safety Act

More recalls by Gk Photonics Inc