RecallDepth

TASE 400 Imaging systems

Company
Collins Aerospace
Recall Initiated
January 29, 2019
Recall Number
Z-1783-2019
Quantity
433 in total
Firm Location
Westford, MA

Reason for Recall

the Nominal Ocular Hazard Distance (NOHD) associated with a Class lllb alignment laser utilized in its TASE family of imaging systems has been miscalculated.

Distribution

US Nationwide Distribution

Lot / Code Info

TASE 400

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Collins Aerospace sent a customer notification letter to affected customers,. The letter identified the affected product problem and actions to be taken. For questions contact Customer Support Team at Support.TASE@collins.com

More recalls by Collins Aerospace