Reprocessed Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheters: Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on GE Systems 10439072RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter Use on Siemens Systems 10438577RH; Biosense Webster SOUNDSTAR eco Diagnostic Ultrasound Catheter, Use on Siemens Systems 10439011RH

Recall Initiated
December 22, 2025
Posted
February 6, 2026
Recall Number
Z-1324-2026
Quantity
94
Firm Location
Northfield, IL

Reason for Recall

These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Distribution

Distribution US nationwide.

Lot / Code Info

10439072RH UDI-DI 10197344157297 Lots EP250724 EP250808; 10439236RH UDI-DI 10197344157303 Lots EP250724 EP250808; 10438577RH UDI-DI 10197344157273 Lots EP250724 EP250808; 10439011RH UDI-DI (case) 20197344157287¿ (ea) 10197344157280¿ Lots EP250724 EP250808

Root Cause

Under Investigation by firm

Action Taken

On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer.