Merit Medical , Custom Manifold Kit REF: K09-13203A
- Company
- Merit Medical Systems, Inc
- Recall Initiated
- December 4, 2025
- Posted
- January 16, 2026
- Recall Number
- Z-1127-2026
- Quantity
- 2192 units
- Firm Location
- Jordan, UT
Reason for Recall
Inflation device handle may detach from the syringe during procedure.
Distribution
Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
Lot / Code Info
K09-13203A: UDI: 00884450479964 /Lot: H2671718 H2681531 H2758436 H2792533 H2796074 H3046979 H3052442 H3147779 H3246149 H3254530 H3297407
Root Cause
Process control
Action Taken
On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days