RecallDepth

Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A

This recall is currently active, issued April 16, 2025. It was issued by Channel Medsystems, Inc..

Company
Channel Medsystems, Inc.
Recall Initiated
March 20, 2025
Posted
April 16, 2025
Recall Number
Z-1584-2025
Quantity
281
Firm Location
Berkeley, CA
Official Source
View on FDA website ↗

Reason for Recall

Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

Distribution

U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.

Lot / Code Info

Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: 108638953, 108645585, 108645586, 108645587, 108866674, 108866675, 109493291, 109493292 and 109895874

Root Cause

Under Investigation by firm

Action Taken

On 03/20/2025, the firm emailed customer an "URGENT MEDICAL DEVICE CORRECTION" Letter informing them that the evaporator gasket inside the handle of the device may not seal properly, which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device. Customers are instructed to: 1. Check your internal inventory to determine if you have any affected devices. 2. Stop use of affected devices. 3. Vent the affected devices by following the attached instructions and then discard the devices according to your facility s procedures. 4. Complete and return the customer acknowledgement form. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also negate the need for us to send any further unnecessary communications on this matter. Therefore, please complete the form even if you no longer have any of the affected devices in your physical inventory. 5. Circulate this letter to all other relevant parties, as needed. For questions, contact Channel Medsystems at safetynotice@cerene.com or at (510)338-9301 Monday thru Friday 8:00 AM to 5:00 PM Pacific Time