RecallDepth

a. Segmental Stem, Cemented, Straight, 100MM, 1 EACH; Model Numbers: CS-09100-03M, CS-10100-03M. b. Segmental Stem, Cemented, Straight, 120MM, 1 EACH. Model Numbers: CS-11120-03M, CS-13120-03M, CS-15120-03M, CS-17120-03M. Component. of ELEOS LIMB SALVAGE SYSTEM

Company
Onkos Surgical, Inc.
Recall Initiated
September 1, 2023
Posted
May 29, 2025
Recall Number
Z-1879-2025
Quantity
147 units
Firm Location
Parsippany, NJ

Reason for Recall

Potential for breach of Tyvek seals.

Distribution

Product was shipped from Onkos' distribution in Olive Branch, MS to consignees in the following states: GA, TX, WA, LA, NJ, FL, AZ, MD, MS, KS, IL, WI, KY, PA, NC, NE, OH, HI.

Lot / Code Info

a. Model Number: CS-09100-03M; UDI-DI: B278CS0910003M0; Lot Numbers: 1880308, 1880309, 1880310, 1880311, 1880312, 1880313, 1880314, 1880315, 1880316, 1880317, 1880318, 1880319, 1880320, 1880321, 1880322, 1880323, 1880324, 1880325, 1880326, 1880327, 1880328, 1880329, 1880330, 1880331, 1880332, 1880333. Model Number: CS-10100-03M. UDI-DI: B278CS1010003M0. Lot Numbers: 1880334, 1880335, 1880336, 1880337, 1880338, 1880339, 1880340, 1880341, 1880342, 1880343, 1880344, 1880345. b. Model Number: CS-11120-03M; UDI-DI: B278CS1112003M0; Lot Numbers: 1880346, 1880347, 1880348, 1880349, 1880350, 1880351, 1880352, 1880353, 1880354, 1880355, 1880356, 1880357, 1880358, 1880359, 1880360, 1880361, 1880362, 1880363, 1880364, 1880365, 1880366, 1880367, 1880368, 1880369, 1880370, 1880371, 1880372, 1880373, 1880374, 1880375, 1880376. Model Number: CS-13120-03M; UDI-DI: B278CS1312003M0; Lot Numbers: 1880377, 1880378, 1880379, 1880380, 1880381, 1880382, 1880383, 1880384, 1880385, 1880386, 1880387, 1880388, 1880389, 1880390, 1880391, 1880392, 1880393, 1880394, 1880395, 1880396, 1880397, 1880398, 1880399, 1880400, 1880401, 1880402, 1880403, 1880404, 1880405, 1880406, 1880407, 1880408, 1880409. Model Number: CS-15120-03M; UDI-DI: B278CS1512003M0; Lot Numbers: 1880410, 1880411, 1880412, 1880413, 1880414, 1880415, 1880416, 1880417, 1880418, 1880419, 1880420, 1880421, 1880422, 1880423, 1880424, 1880425. Model Number: CS-17120-03M; UDI-DI: B278CS1712003M0; Lot Numbers: 1880426, 1880427, 1880428, 1880429, 1880430, 1880431, 1880432, 1880433, 1880434, 1880435, 1880436, 1880437, 1880438, 1880439, 1880440, 1880441, 1880442, 1880443, 1880444, 1880445, 1880446, 1880447, 1880448, 1880449, 1880450, 1880451, 1880452, 1880453, 1880454.

Root Cause

Packaging change control

Action Taken

Onkos Surgical began notifying consignees on about 09/01/2023 via email. Consignees were instructed to complete and return the response form providing the affected lot numbers being returned, coordinate the return of affected units, and to note that when placing these parts into totes, always place them flat.

More recalls by Onkos Surgical, Inc.

View all recalls by this company →