RecallDepth

EMPOWR 3D Knee Tibial Insert, 5R 16MM, VE, REF: 342-16-705; EMPOWR 3D Knee Tibial Insert, 3R 14MM, VE, REF: 342-14-703

This recall is currently active, issued December 26, 2024.

Recall Initiated
December 18, 2024
Posted
December 26, 2024
Recall Number
Z-0766-2025
Quantity
33
Firm Location
Austin, TX
Official Source
View on FDA website ↗

Reason for Recall

5R 16MM knee tibial insert package may contain knee tibial insert 3R 14MM and vice versa, which may cause surgical delay to get a new product.

Distribution

US Nationwide distribution in the states of VA, MN, TX, WI, IA, IL, IN, SC, NY, MD, CA, FL, AZ, WA.

Lot / Code Info

REF/UDI-DI/Lot: 342-16-705/00888912167628/159T1086 342-14-703/00888912167505/137T1086

Root Cause

Labeling Change Control

Action Taken

On 12/18/2024, recall notices were emailed to consignees who were asked to do the following: 1. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 2. Review your stock for the products and lot numbers for the items to be returned. 3. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service. 4. Complete and return the acknowledgement and response form via email to Lesli.Helmick@enovis.com If you have questions contact the firm at productsafety@enovis.com