Altair 330 lbs w/o Hanger Bar, REF: 55200H-NA, 55220H-NA, and Roomer S 220 kg w/o Hanger Bar, REF: 55300H-NA, patient lifts
This recall is currently active, issued April 5, 2024. It was issued by Human Care Hc Sweden Ab Arstaangsvagen 21 B Stockholm Sweden.
- Recall Initiated
- February 23, 2024
- Posted
- April 5, 2024
- Recall Number
- Z-1486-2024
- Quantity
- 10
- Official Source
- View on FDA website ↗
Reason for Recall
Patient lift hanger bar may come loose, which could potentially pose risks to patients.
Distribution
US Nationwide distribution in the states of TX, IL, OK, CA.
Lot / Code Info
UDI-DI: 7331599999943 Products manufactured between May 2023 and October 2023. Serial Numbers: 236229060002, 236229060005, 236232010015, 236229070004, 236232010006, 236229060003, 236229060009, 236229060006, 236232010005, 236232010003
Root Cause
Process control
Action Taken
On 2/23/24, Field Safety Notices were emailed to customers who were asked to do the following: 1) Checking the fixation by carefully pulling on the shaft. Should you observe any deviation or irregularity, please inform us promptly. 2) Read and understand this letter and provide a confirmation via email to the address quality@humancaregroup.com In case of recognizing any deviation or irregularity during the fixation control or if in need of additional information email quality@humancaregroup.com On 5/6/24, Field Safety Notices were emailed to customers with additional instructions for distributors and dealers to distribute the notice to all customers who have purchased the impacted products.