RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
This recall is currently active, issued October 31, 2023. It was issued by Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland.
- Recall Initiated
- September 6, 2023
- Posted
- October 31, 2023
- Recall Number
- Z-0200-2024
- Quantity
- 35 units
- Official Source
- View on FDA website ↗
Reason for Recall
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
Distribution
US Nationwide distribution including Puerto Rico
Lot / Code Info
GTIN: 05055273206104 Lot/Batch Number: (1) 7201-0747 7201-0423 7201-0417 7201-0353 7255-0198 7255-0172 7255-0171 7255-0129 7201-1085 7201-1086 7201-1079 7201-1080 7201-0982 7201-0977 7201-0970 7201-0950 7201-0924 7201-0919 7201-0901 7201-0849 7201-0827 7201-0801 7201-0768 7201-0606 7201-0562 7201-0535 7201-0531 7201-0482 (2)¿¿¿¿¿¿¿¿¿¿¿¿¿ 7201-0776 7201-0367 7201-0343¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿ 7201-0348 ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿
Root Cause
Under Investigation by firm
Action Taken
Randox issued Medical Device Correction letter (REC687) on 9/13/23 to to US and PR customers on 9/12/23. Letter states reason for recall, health risk and action to take: Randox recommendations: " Auto rerun settings are disabled, see Appendix A for guidance for each Rx instrument. " Results with flag(<) reported as the bottom of the technical range. Sample Volume settings Low should not be applied. " Results with flag (>) reported as the top of the technical range. Sample Volume settings high and dilutions should not be applied. A manual dilution should be performed, and results multiplied by the dilution factor. The RX Parameter rerun settings will be updated with a parameter release in due course, the Instructions For Use will also be updated at this time. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.