Medical Device Identification Cards associated with Sprint Quattro Lead, Product/CFN 6935M72

Company
Medtronic Inc., Cardiac Rhythm And Heart Failure (Crhf)
Recall Initiated
November 8, 2021
Recall Number
Z-0390-2022
Quantity
187 Medical Device Cards
Firm Location
Mounds View, MN

Reason for Recall

The back of some Medical Device Identification Cards provided to patients were printed with an erroneous statement stating the following: "This patient has a complete MR conditional system implanted."

Distribution

US, Canada, Northern Mariana Islands

Lot / Code Info

Sprint Quattro Lead, Product/CFN 6935M72,  UDI 00643169550919, Serial Numbers: TDL440798V, TDL475850V, TDL473645V, TDL386490V, TDL446689V, TDL458696V, TDL398634V, TDL440035V, TDL458694V, TDL397987V, TDL469404V, TDL469730V, TDL469411V, TDL473637V, TDL434492V, TDL433920V, TDL469403V, TDL397982V, TDL448256V, TDL398632V, TDL424268V, TDL448531V, TDL458699V, TDL460247V, TDL410929V, TDL405398V, TDL440809V, TDL448519V, TDL410295V, TDL410332V, TDL445702V, TDL368164V, TDL378146V, TDL446690V, TDL386481V, TDL458695V, TDL448518V, TDL448258V, TDL433896V, TDL446674V, TDL397974V, TDL386494V, TDL429017V, TDL397966V, TDL369718V, TDL424266V, TDL445620V, TDL445705V, TDL440984V, TDL440027V, TDL386489V, TDL403279V, TDL405769V, TDL434493V, TDL440803V, TDL445618V, TDL448527V, TDL433902V, TDL440971V, TDL440977V, TDL393417V, TDL433899V, TDL445712V, TDL403276V, TDL410915V, TDL433921V, TDL410334V, TDL414801V, TDL395189V, TDL429015V, TDL410291V, TDL410296V, TDL445703V, TDL368162V, TDL397975V, TDL393407V, TDL401828V, TDL423546V, TDL410331V, TDL431773V, TDL345939V, TDL350436V, TDL410916V, TDL368172V, TDL386480V, TDL440026V, TDL386477V, TDL410328V, TDL410301V, TDL410912V, TDL397992V, TDL344964V, TDL397985V, TDL431745V, TDL349616V, TDL405394V, TDL369722V, TDL433898V, TDL433909V, TDL424273V, TDL393401V, TDL423555V, TDL424267V, TDL397955V, TDL410905V, TDL349118V, TDL423556V, TDL350442V, TDL386479V, TDL368161V, TDL431777V, TDL431778V, TDL330398V, TDL381963V, TDL326160V, TDL345936V, TDL414787V, TDL386499V, TDL424269V, TDL386495V, TDL397967V, TDL405399V, TDL327272V, TDL405767V, TDL386493V, TDL315797V, TDL414793V, TDL386486V, TDL410341V, TDL386487V, TDL361821V, TDL386476V, TDL397983V, TDL312450V, TDL395193V, TDL378138V, TDL393408V, TDL337172V, TDL321443V, TDL395190V, TDL386496V, TDL397977V, TDL326159V, TDL369709V, TDL386497V, TDL369720V, TDL397958V, TDL298585V, TDL321456V, TDL386483V, TDL369713V, TDL397973V, TDL386498V, TDL386485V, TDL349115V, TDL336148V, TDL393409V, TDL344961V, TDL350065V, TDL337169V, TDL393405V, TDL349559V, TDL343795V, TDL393400V, TDL294118V, TDL393403V, TDL316186V, TDL368175V, TDL337170V, TDL326110V, TDL316181V, TDL323585V, TDL321171V, TDL315812V, TDL316169V, TDL276334V, TDL298584V, TDL199017V  UDI 00613994718877, Serial Numbers: TDL098796V, TDL046363V, TDL016560V, TDL016558V, TDL016557V, TDL002704V, TDL010685V  UDI 00643169309784, Serial Numbers: TDL192485V, TDL187462V

Root Cause

Error in labeling

Action Taken

On 09-Nov-2021, Medtronic initiated mailing the patient letter, including a corrected Medical Device Identification Card, to impacted patients. In addition, on 08-Nov-2021, physicians following impacted patients were sent a physician letter explaining the correction.