Philips Incisive CT -Computed Tomography X-Ray System Model Number: 72813
- Company
- Philips North America
- Recall Initiated
- March 20, 2023
- Posted
- May 12, 2023
- Recall Number
- Z-1586-2023
- Quantity
- 3 units US ;2 units OUS
- Firm Location
- Cambridge, MA
Reason for Recall
Multi-Function Foot Switch Unload Pedal Foot Entrapment-Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operator's foot and injury
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, MI, MO, NY, TN, TX, VT and the countries of Australia, Chile, Costa Rica, Czech Republic, Dominican, Republic Germany, India, Israel, Japan, Lithuania, Philippines, Poland, Sint Maarten (Dutch part), South Africa, Thailand, United Kingdom, Utd. Arab. Emir.
Lot / Code Info
UDI-DI:(01)00884838085015(21) + Serial number US Serial Numbers: 33041, 500301, 500252. OUS Serial Numbers: 33040, 500408
Root Cause
Under Investigation by firm
Action Taken
Philips Medical issued Urgent Medical Device Correction Letter Multi-Function Switch Unload Pedal Foot Entrapment on 3/20/23 to US consignees via Certified United States Postal Service (USPS) with delivery confirmation. International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Immediately stop using the Unload Pedal of the Foot Switch until Philips installs a solution on your system. " Instead of using the Unload Pedal, please use the Unload Patient function on the Gantry Control Panels, or use other Table In/Out, Up/Down functions on the Gantry Control Panels, CT Scan Control Box, or Hardware lnterventional Controls to fulfill the normal patient unloading function, as instructed in the IFU. " You may continue to use Load Pedal, and Free-Float Pedal of Foot Switch, as they are not affected by this Urgent Medical Device Correction Letter. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the solution to resolve this issue (refer to FCO72800794). If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).