ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RMC

Company
Cordis Us Corp
Recall Initiated
March 24, 2023
Posted
May 11, 2023
Recall Number
Z-1485-2023
Quantity
471 units
Firm Location
Miami Lakes, FL

Reason for Recall

There is a potential for separation of the ANGIOGUARD RX / XP delivery system and capture sheath which include but are not limited to situations of an intra-procedural delay, unplanned percutaneous or surgical intervention, or stroke; while a replacement device is prepared.

Distribution

Worldwide distribution.

Lot / Code Info

UDI/DI 20705032056745; Lot Numbers: 35263328, 35264207, 35264213, 35264216, 35265393, 35265648, 35265659

Root Cause

Vendor change control

Action Taken

The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.

More recalls by Cordis Us Corp

View all recalls by this company →