RecallDepth

VITROS Chemistry Products VersaTips, Catalog No. 6801715 A VersaTip sales unit is composed of 1000 tips per box.

Company
Ortho-Clinical Diagnostics, Inc.
Recall Initiated
December 10, 2021
Recall Number
Z-0735-2022
Quantity
4125 units
Firm Location
Rochester, NY

Reason for Recall

The lot has a shape deformity which may lead to a 16% reduction in the volume of the sample or reagent dispensed. In cases where the error is not detected and the affected results are not suppressed, clinical impact may include reduced drug dosage, or false positive or false negative results for bloodborne infectious disease assays when sample results are close to the cut-off value.

Distribution

Domestic distribution nationwide. Foreign distribution to Canada.

Lot / Code Info

UDI 10758750006786  Lot 1039734

Root Cause

Nonconforming Material/Component

Action Taken

On December 10, 2021, the firm notified consignees through ORTHO PLUSSM (a service that provides customers with Ortho Clinical Diagnostics product notifications electronically) of the product issue. Customers were instructed to discontinue use and discard the affected lot. Customers should forward the notification if the product was distributed outside of their facility.