Signa 1.5T TwinSpeed Magnetic Resonance Imaging System

Company
Ge Healthcare, Llc
Recall Initiated
February 18, 2022
Posted
April 4, 2022
Recall Number
Z-0857-2022
Quantity
10 units
Firm Location
Waukesha, WI

Reason for Recall

There is potential for the images to be flipped left to right.

Distribution

Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.

Lot / Code Info

System IDs: MMIMR01, 213748HDMR, 714125MR, 561218MR1, 918523MR1,304766MR1, 082427030026, 082427080007, 082427020017, 376569401MR

Root Cause

Device Design

Action Taken

This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.

More recalls by Ge Healthcare, Llc

View all recalls by this company →