ACUSON Juniper Diagnostic Ultrasound System

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
July 12, 2021
Terminated
October 16, 2024
Recall Number
Z-2471-2021
Quantity
15 systems
Firm Location
Issaquah, WA

Reason for Recall

Due to intermittent failures of the power supply in the ultrasound system which renders it inoperable.

Distribution

Worldwide distribution - U.S. Nationwide distribution in the Puerto Rico and the countries of Bosnia Herzegovina, Canada, Colombia, Germany, Greece, Italy, North Macedonia, Poland, and Slovenia.

Lot / Code Info

Model Number: 11335791 UDI: (01)04056869152059 (21)123456 (240)11335791 (422)410 Firmware version 3.8 for rev. 7 U.S. Serial Number: 555606; O.U.S. Serial Numbers: 550975; 552770; 553629; 553951; 555060; 555601; 555602; 555604; 555605; 555610; 555611; 555612; 555613; 555626; O.U.S. Inactive Serial Numbers: 555603; 555607;

Root Cause

Software change control

Action Taken

On 07/12/2021, the firm sent a "MEDICAL DEVICE SAFETY CORRECTION" Notification via email to customers informing them that there are intermittent failures of the power supply in the ultrasound system which may render the system inoperable. When an ultrasound system experiences this failure, cycling the power does not recover system functionality. Potential health risks include, lack of ultrasound guidance during interventional procedures (e.g., biopsy, guidance during surgery, patient under sedation) that rely on real-time ultrasound imaging to guide or complete the procedure. Customer will be contacted by a Customer Service Engineer to schedule a visit to their facility to update the system. Customers are asked to ensure that all users are aware of this issue and correction within their organization and others who may need to be inform of this situation. For further questions, contact Siemens Healthineers USD Complaints at usd-complaint.team@siemens-healthineers.com