DELTAVEN Y, Safety I.V. Catheter in Pur with closed system a) REF 3832072 b) REF 3810072 c) REF 3842072 d) REF 3852072
- Recall Initiated
- June 21, 2021
- Recall Number
- Z-2298-2021
Reason for Recall
Problems related to the sterilization of the medical devices; possible missing sterility of the medical device
Distribution
US Nationwide distribution in the states of FL and MN.
Lot / Code Info
a) REF 3832072, Lot 11T80190 b) REF 3810072, Lot 11T80303 c) REF 3842072, Lot 12T80342 d) REF 3852072, Lot 12T80348
Root Cause
Process control
Action Taken
The firm disseminated the notices on 06/21/2021 by email. The firm is requesting destruction of the affected devices at the consignee.