RecallDepth

Medicina ENFit 5ml Reusable Enteral Syringe Code: LHE05

Company
Medicina Uk Ltd
Recall Initiated
July 4, 2022
Posted
October 20, 2022
Recall Number
Z-0108-2023
Quantity
1 box (100 syringes)

Reason for Recall

Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification

Distribution

Nationwide

Lot / Code Info

UDI-DI: (01)05060278508153(17)250400(10)HM20055 Batch number: HM17099, HM18010, HM18052, HM18103, HM18113, HM18143, HM19009, HM19018, HM19061, HM19084, HM19132, HM19146, HM20006, HM20032, HM20055, HM20095, HM21037, HM21069, HM22019

Root Cause

Nonconforming Material/Component

Action Taken

Medicina Ltd issued Field Safety Corrective Action notice to the US Distributor 07/04/2022 via email. The US Distributor is conducting recall to the user level letter issued the week of July 25th via United States Postal Service. . Letter states reason for recall, health risk and action to take: Action to be taken - Please carefully read this Field Safety Notice and return the Acknowledgement Form - Please do not use any of the batches of product listed on this notice - If you have product to return, please complete the information on the Acknowledgement Form and our team will arrange the return for you - Please distribute throughout your organisations and to relevant personnel - If you have any questions or queries in relation to this notice, please contact us using the information below Contact information regarding FSN Shaista Patel QA/RA Administrator Unit 2, Rivington View Business Park, Station Road, Blackrod, Bolton BL6 5BN Email: quality.administrator@medicina.co.uk Tel: 01204 695050

More recalls by Medicina Uk Ltd