Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)

Company
Percussionaire Corporation
Recall Initiated
March 28, 2023
Posted
January 17, 2024
Recall Number
Z-0693-2024
Quantity
195
Firm Location
Sandpoint, ID

Reason for Recall

Breathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.

Distribution

US: RI, CA, TX

Lot / Code Info

UDI-DI: 00849436000693, Lots/Expiration: WO03739/September 13, 2024, and WO04218/November 14, 2024.

Root Cause

Process control

Action Taken

On 3/28/23, correction notices were emailed to customers who were asked to remove and switch the reservoir bag to the proper location. On 1/24/24, follow-up correction notices were sent to customers who were asked to do the following: 1) Review your inventory, and if affected devices are identified, ensure the reservoir bag is in the proper location. 2) If affected devices were further distributed, send a copy of this notice to your customers. 3) Complete and return the Effectiveness Check Questionnaire via email to customerservice.us@sentec.com If you have any questions, contact the firm's Customer Service at customerservice.us@sentec.com or by phone at (208)- 263- 2549 Monday - Friday from 7 AM to 3:30 PM (PST). Additional information for contact information can be found at www.sentec.com/contact