CombiDiagnost R90 1.1 (709031)

Company
Philips Medical Systems
Recall Initiated
February 11, 2022
Posted
February 14, 2024
Recall Number
Z-1088-2024
Quantity
130 units
Firm Location
Cambridge, MA

Reason for Recall

While performing a fluoroscopy examination, there is a potential that the Radio Fluoroscopy (RF) viewer will also display a previous patient's radiography images. If the issue occurs, there will be differences in image content, image format and image size.

Distribution

US Nationwide distribution in the states of CA, FL, GA, IL, IN, KS, MD, ME, MI, MN, NC, NJ, NY, TX, VA, WA.

Lot / Code Info

Model: R90 1.1 (709031); UDI DI: 00884838101456; Software version 1.1.1 or higher; Serial Numbers: 10001004, 10001005, 10001006, 10001008, 10001009, 10001010, 10001011, 10001012, 10001013, 10001018, 10001020, 10001022, 10001024, 10001026, 10001027, 10001028, 10001029, 10001030, 10001031, 10001032, 10001035, 10001036, 10001043, 10001044, 10001045, 10001046, 10001047, 10001048, 10001049, 10001050, 10001051, 10001052, 10001053, 10001054, 10001055, 10001056, 10001057, 10001058, 10001059, 10001060, 10001061, 10001062, 10001063, 10001064, 10001065, 10001066, 10001067, 10001068, 10001069, 10001070, 10001071, 10001072, 10001073, 10001074, 10001075, 10001076, 10001077, 10001078, 10001079, 10001080, 10001081, 10001083, 10001084, 10001085, 10001086, 10001087, 10001088, 10001089, 10001090, 10001091, 10001092, 10001093, 10001094, 10001095, 10001096, 10001097, 10001098, 10001099, 10001100, 10001101, 10001102, 10001103, 10001104, 10001105, 10001106, 10001107, 10001108, 10001109, 10001110, 10001111, 10001112, 10001113, 10001114, 10001115, 10001116, 10001117, 10001118, 10001119, 10001120, 10001121, 10001122, 10001123, 10001124, 10001125, 10001126, 10001127, 10001128, 10001129, 10001130, 10001131, 10001132, 10001133, 10001134, 10001135, 10001136, 10001137, 10001138, 10001139, 10001140, 10001141, 10001142, 10001143, 10001144, 10001145, 10001147, 10001148, 10001149, 10001150, 10001152, 10001201;

Root Cause

Software design

Action Taken

An Important Product Notice (2021-PD-DXR-034) dated February 15, 2022 was issued via Certified Mail. Philips Field Service Engineers will contact customers to schedule an onsite visit to install update software correcting the issue. Customers are asked to follow the provided instructions to ensure there are no pending images before an exam circulate the notice to all users and to complete and return the response form and circulate the notice to all users of the device for awareness. Contact Philips Customer Care Solutions Center 1-800-722-9377 with questions.