Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2

Company
Beckman Coulter Inc.
Recall Initiated
July 12, 2023
Posted
August 31, 2023
Recall Number
Z-2488-2023
Quantity
12,850 units
Firm Location
Brea, CA

Reason for Recall

Lipemic interference failed to meet the performance specification listed within the IFU.

Distribution

US: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE NJ NM NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV OUS: Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, C¿te d'Ivoire, Croatia, Cura¿ao, Cyprus, Czechia, Denmark, Ecuador, Egypt, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uruguay, Viet Nam, Yemen.

Lot / Code Info

All Lots/UDI-(01)15099590011574

Root Cause

Other

Action Taken

On July 18, 2023 Beckman Coulter issued a "Urgent Medical Device Recall" Notification via email and postal mail. Beckman Coulter asked consignees to take the following action: 1. No retrospective re view of past patient results is required however Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Discontinuance or disposal of this product is not necessary. 3. Per the IFU, avoid highly lipemic samples when using the IgA assay. 4. No update is required to the LIH influences check settings on the analyzer, where in use. 5. Please share this information with your laboratory staff nd retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 6. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: - Electronically - If you received this communication via email, Manually - complete and return the enclosed Response Form. 7. If you have any questions regarding this notice, please contact our Customer Support Center - www.beckmancoulter.com or 1-800-854-3633 8. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by Beckman Coulter Inc.

View all recalls by this company →