Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53* 0998-00-0800-55* *Not US distribution

Company
Datascope Corp.
Recall Initiated
June 5, 2023
Posted
July 20, 2023
Recall Number
Z-2151-2023
Quantity
9174 (4586 US; 4588 OUS) in total
Firm Location
Mahwah, NJ

Reason for Recall

An unexpected shutdown of the IABP may occur due to failures of tantalum capacitors in the Power Management Board and/or Solenoid Board. This issue may lead to an unexpected interruption of therapy.

Distribution

Worldwide Distribution- US (Nationwide) and OUS (foreign) countries of: ALBANIA, ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, AZERBAIJAN , BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GHANA, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM, YEMEN, ZAMBIA

Lot / Code Info

All Unit Serial Numbers Model UDI 0998-00-0800-31 10607567109053 0998-00-0800-32 10607567111117 0998-00-0800-33 10607567109008 0998-00-0800-34 10607567111940 0998-00-0800-35 10607567109107 0998-00-0800-45 10607567108421 0998-00-0800-52 10607567108438 0998-00-0800-53 10607567108391 0998-00-0800-55 10607567108414 0998-00-0800-65 10607567113432

Root Cause

Component design/selection

Action Taken

On June 5, 2023, the firm notified customers of the recall via "URGENT MEDICAL DEVICE CORRECTION" letters. Letters covered two issues: Issue 1 - Failure to charge battery (Covered by RES 92287) Issue 2 - Unexpected shutdown (Covered by this RES event) User Actions: Issue 2: Unexpected Shutdown due to failure of Tantalum Capacitors. Should loss of power occur, therapy will be interrupted and the console will need to be exchanged with another (if available) to resume therapy. If your device remains inoperable, please contact your service representative. It is recommended for users to have an alternate IABP on hand in the event that this issue may occur. The firm is developing a hardware correction to address this issue. If you have any questions, please contact your Datascope/Getinge representative or call Datascope/Getinge Technical Support at 1-888-943-8872, options 4, 2, 1, Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. ET