Power-PRO 2 . Intended as an ambulance cot to transport a patient to or from an emergency or non-emergency location. Model: 650705550001

Company
Stryker Medical Division Of Stryker Corporation
Recall Initiated
February 1, 2023
Posted
March 15, 2023
Recall Number
Z-1256-2023
Quantity
4 units
Firm Location
Portage, MI

Reason for Recall

Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.

Distribution

US Nationwide distribution in the states of NY, VA.

Lot / Code Info

GTIN: 07613327559118 Serial Numbers: 2208002066, 2208002071, 2209001508, 2208001664

Root Cause

Under Investigation by firm

Action Taken

Stryker issued Urgent Medical Device Correction letter dated 2/1/23. Letter states reason for recall, health risk and action to take: 1. Locate the units listed on the attached business reply form and remove them from service until repairs are complete. 2. Return the enclosed business reply form to confirm receipt of this notification by fax +1 269 488 8691 or email productfieldaction@stryker.com. 3. Stryker will contact you to arrange for the modification/repair of your Power-PRO 2. 4. If you have loaned or sold or disposed of any of the products listed in this letter, please forward a copy of this notice to the new users and advise us of their new location in the space provided on the business reply form. Questions or concerns, contact Customer Service at +1 800 327 0770 Monday through Friday from 8:00 a.m. to 6:00 p.m. ET.

More recalls by Stryker Medical Division Of Stryker Corporation

View all recalls by this company →