Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.
This recall is currently active, issued August 23, 2024. It was issued by Treace Medical Concepts, Inc..
- Company
- Treace Medical Concepts, Inc.
- Recall Initiated
- June 2, 2024
- Posted
- August 23, 2024
- Recall Number
- Z-2718-2024
- Quantity
- 40 kits containing 40 plates
- Firm Location
- Ponte Vedra, FL
- Official Source
- View on FDA website ↗
Reason for Recall
Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.
Distribution
US Nationwide distribution in the states of CO, FL, MD, MI, NC, SC, TX, VA.
Lot / Code Info
UDI: (01)00810111222464(17)290424(10)300542404
Root Cause
Under Investigation by firm
Action Taken
Distributors were contacted via phone or in person to hold product starting on 06/02/2024. Recall notifications were disseminated on 06/13/2024 via email containing the recall letter, acknowledgement, and shipping label for return. On 07/29/2024, Treace Medical contacted surgeons via letter. Surgeons were instructed to monitor patients for adverse events and to complete and return the acknowledgement form.