INNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cup

Company
Innomed, Inc.
Recall Initiated
March 7, 2024
Posted
May 1, 2024
Recall Number
Z-1753-2024
Quantity
41 devices
Firm Location
Savannah, GA

Reason for Recall

54 MM finisher blade was inadvertently affixed to the shaft of the CupX handle in lieu of the 54 MM starter blade.

Distribution

FL, GA, PA, NY, CA, VA, SD, MN, TX, UT, NC, and UK, Denmark, Germany Switzerland, Ireland, Canada

Lot / Code Info

UDI/DI 00840277104007, Lot Number 1223K

Root Cause

Under Investigation by firm

Action Taken

Innomed issued an urgent medical device recall notice to its consignees on 03/07/2024 via email. The notice explained the issue, potential risk to health, and requested the return of the affected devices for US distribution and destruction of the affected devices distributed outside the US.