RecallDepth

T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300

This recall has been terminated (originally issued April 15, 2021).

Company
Flower Orthopedics Corporation
Recall Initiated
April 15, 2021
Terminated
November 12, 2021
Recall Number
Z-1698-2021
Quantity
47 units
Firm Location
Horsham, PA
Official Source
View on FDA website ↗

Reason for Recall

Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages

Distribution

US Nationwide distribution in the states of AZ, CO, FL, GA, IL, IN, PA, NC, VA.

Lot / Code Info

Lot Numbers: 2003320898, 1810220041, 1802220008, 1802220005, 1711220035  UDI: 00840118110853

Root Cause

Nonconforming Material/Component

Action Taken

Flower Orthopedic issued Urgent Recall Information via email on 4/15/21. Letter states reason for recall, health risk and action to take: Check your inventory for the product listed below and quarantine immediately. Next, please complete and return the attached Acknowledgement and Receipt Form as soon as possible. Flower Orthopedics is initiating shipment of replacement product for delivery Friday April 16. Contact information for questions: Monday through Friday, 8:30 AM to 5:00 PM, EDST 877-778-8587 or FLOWERCS@FLOWERORTHO.COM "Acknowledgement and Product Replacement Forms (see attached)