Braive Growth Modulation System Break-Off Set Screws - Product Usage: intended to limit further progression of the spinal curvature and provide correction of the thoracic curve while allowing continued skeletal growth by compression of the convex side of the scoliotic curve.

Company
Medtronic Sofamor Danek Usa Inc
Recall Initiated
April 27, 2021
Terminated
August 20, 2024
Recall Number
Z-1797-2021
Quantity
162 UNITS
Firm Location
Memphis, TN

Reason for Recall

Product is impacted by a thread profile defect due to a manufacturing issue.

Distribution

International distribution in the countries of France, Switzerland and Jordan. No US distribution.

Lot / Code Info

CT20L012 and CT20L020.  CT20L012 EXP 2 DEC 2028 CT20L020 EXP 11 DEC 2028

Root Cause

Under Investigation by firm

Action Taken

On 4/26/2021 the firm sent a letter to its consignees with the following instructions: Required Actions: 1) Identify, segregate, and quarantine any impacted product (listed above) within your inventory. 2) Return the impacted product to Medtronic. Your Medtronic Sales Representative can assist in facilitating the return of product as necessary. 3) Please maintain a copy of this notice in your records. 4) Please share this communication within your organization, with other organizations where affected devices have been transferred, and any other associated organizations that may be impacted by this action.