Medtronic HVAD Monitor Data Cable, Product Number: 1575

Company
Heartware, Inc.
Recall Initiated
February 26, 2021
Recall Number
Z-1339-2021
Quantity
1337 units (1110 US)
Firm Location
Miami Lakes, FL

Reason for Recall

There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.

Distribution

Worldwide

Lot / Code Info

All lot codes

Root Cause

Device Design

Action Taken

The firm issued a Medical Device Notice to its consignees on 02/26/2021 by letter delivered via UPS 2-day delivery. The letter explained the issue and directed the physician to review the patient management recommendations on the frequency of of inspecting the HVAD Controller, HVAD Controller power cables, and HVAD Monitor data cables. For questions regarding this material, please contact your Medtronic Field Representative.