Aestiva/5 MRI Anesthesia System - Product Usage: The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss.
- Company
- Ge Healthcare, Llc
- Recall Initiated
- September 17, 2020
- Recall Number
- Z-0311-2021
- Quantity
- 2 devices
- Firm Location
- Waukesha, WI
Reason for Recall
Flow sensors could have damaged tubes in the form of small punctures or cuts.
Distribution
U.S. states where product was distributed - AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI Foreign countries impacted - Australia, Canada, Czech Republic, Estonia, France, Germany, India, Indonesia, Israel, Italy, Kuwait, Latvia, Macedonia, Madagascar, Malaysia, Nicaragua, Panama, Qatar, Russia, Singapore, Slovenia, Spain, Thailand, Ukraine, United Kingdom
Lot / Code Info
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)
Root Cause
Process control
Action Taken
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.