Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

Company
Ge Healthcare, Llc
Recall Initiated
December 10, 2020
Recall Number
Z-0856-2021
Quantity
81 units
Firm Location
Waukesha, WI

Reason for Recall

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Distribution

Worldwide

Lot / Code Info

Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110,  63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY50108, 63GY50121, 63GZ50142, 63GY50112, 63GZ50139, 63GY50107, 63GZ50138   NM 830 (Model Number H3910AC), Serial Numbers: 830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186   830 NM Cyber Defense (Model Number H3910AG), Serial Numbers: 830Z60185

Root Cause

Process control

Action Taken

The firm initiated their action by letter to the consignees on 12/10/2020. The letter explained the issue and stated that GE would inspect, and if required, correct the affected units. Questions and concerns may be directed to GE Healthcare Service at 1-800-437-1171 or the consignee's local Service Representative.