VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip

This recall has been terminated (originally issued July 27, 2017).

Company
Medtronic Navigation, Inc.
Recall Initiated
July 27, 2017
Terminated
August 5, 2019
Recall Number
Z-0025-2018
Quantity
423 devices
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.

Distribution

US nationwide distribution.

Lot / Code Info

Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip Lots - 000212782059, 000212823919, 000212894256, 000212978141, 000212783784, 000212854585, 000212926008, 000212996201, 000212823883, 000212882619, 000212956163, 000213145362; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip Lots - 000212354781, 000212656015, 000212760821, 000212913269, 000212384009, 000212660280, 000212827856, 000212922387, 000212465594, 000212664497, 000212838837, 000212935278, 000212507737, 000212698642, 000212846727, 000213000097, 000212510629, 000212715926, 000212856596, 000213001255, 000212624409, 000212745633, 000212873863, 000213009483, 000212626439, 000212753561, 000212873873, 000213041517, 000212632885, 000212756095, 000212879480, 000213056030; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip Lots - 000212141557, 000212503634, 000212647830, 000212796559, 000212148368, 000212504463, 000212701679, 000212846726, 000212434500, 000212549456, 000212707035, 000212901580, 000212438405, 000212553428, 000212745660, 000212912501, 000212473141, 000212632884, 000212795235, 000212963182

Root Cause

Process design

Action Taken

All US consignees will be a sent a customer letter, dated July 27, 2017, via FedEx or Email. In addition, customers will be asked to return a signed consignee response form.