VCLAS (Visualase Cooled Laser Ablation System) Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip
This recall has been terminated (originally issued July 27, 2017).
- Company
- Medtronic Navigation, Inc.
- Recall Initiated
- July 27, 2017
- Terminated
- August 5, 2019
- Recall Number
- Z-0025-2018
- Quantity
- 423 devices
- Firm Location
- Louisville, CO
- Official Source
- View on FDA website ↗
Reason for Recall
The VCLAS 3mm, 10mm, and 15mm tip devices were not appropriately tested for saline leakage during the manufacturing process.
Distribution
US nationwide distribution.
Lot / Code Info
Part Number: 9735559 Visualase Cooled Laser Applicator System 3mm Tip Lots - 000212782059, 000212823919, 000212894256, 000212978141, 000212783784, 000212854585, 000212926008, 000212996201, 000212823883, 000212882619, 000212956163, 000213145362; Part Number: 9735560 Visualase Cooled Laser Applicator System 10mm Tip Lots - 000212354781, 000212656015, 000212760821, 000212913269, 000212384009, 000212660280, 000212827856, 000212922387, 000212465594, 000212664497, 000212838837, 000212935278, 000212507737, 000212698642, 000212846727, 000213000097, 000212510629, 000212715926, 000212856596, 000213001255, 000212624409, 000212745633, 000212873863, 000213009483, 000212626439, 000212753561, 000212873873, 000213041517, 000212632885, 000212756095, 000212879480, 000213056030; Part Number: 9735561 Visualase Cooled Laser Applicator System 15mm Tip Lots - 000212141557, 000212503634, 000212647830, 000212796559, 000212148368, 000212504463, 000212701679, 000212846726, 000212434500, 000212549456, 000212707035, 000212901580, 000212438405, 000212553428, 000212745660, 000212912501, 000212473141, 000212632884, 000212795235, 000212963182
Root Cause
Process design
Action Taken
All US consignees will be a sent a customer letter, dated July 27, 2017, via FedEx or Email. In addition, customers will be asked to return a signed consignee response form.