STA - Neoplastine Cl (ref. 00605) Product Usage: Manual or automated determination of the prothrombin time (PT).

Company
Diagnostica Stago, Inc.
Recall Initiated
October 25, 2017
Terminated
May 12, 2020
Recall Number
Z-1068-2018
Firm Location
Parsippany, NJ

Reason for Recall

Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).

Distribution

Nationally

Lot / Code Info

Lot 251064 (Exp: 09-2018) Lot 251725 (Exp: 02-2019)

Root Cause

Device Design

Action Taken

Stago sent an Urgent Field Safety Notice letter dated October 25, 2017 to their customers. The letter identified the affected product, problem and actions to be taken: Customers were instructed to: If it is not already done, run a Quality Control test at every change of vial (if you have affected lots of product). Return to company, by fax or email, an acknowledgement form confirming that the recall letter was read by the receiving company. For questions contact the Stago Hotline (1-800-725-0607).