VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents
This recall has been terminated (originally issued March 11, 2019).
- Company
- Ortho-Clinical Diagnostics
- Recall Initiated
- March 11, 2019
- Terminated
- December 7, 2020
- Recall Number
- Z-1222-2019
- Quantity
- 80
- Firm Location
- Rochester, NY
- Official Source
- View on FDA website ↗
Reason for Recall
One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.
Distribution
Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom
Lot / Code Info
Systems Running Software Version 3.4 & 3.4.1
Root Cause
Software design
Action Taken
Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.