VITROS XT 7600 Integrated System, Product Code 6844461, UDI 1075870031658 Product Usage: The VITROS XT 7600 Integrated System is intended for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents, and VITROS Immunodiagnostic Products Reagents

This recall has been terminated (originally issued March 11, 2019).

Company
Ortho-Clinical Diagnostics
Recall Initiated
March 11, 2019
Terminated
December 7, 2020
Recall Number
Z-1222-2019
Quantity
80
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

One of the software algorithms used to detect sample dispense errors was inadvertently disabled. Because of this, sample dispense errors may lead to incorrect results being reported without an error code to alert the user.

Distribution

Worldwide Distribution - US Nationwide in the states of: AZ, CA, FL, IL, OH, TX, and VA. The products were distributed to the following foreign countries: Australia, Belgium, Canada, Chile, Columbia, France, India, Italy, Japan, Portugal, Singapore, Spain, United Kingdom

Lot / Code Info

Systems Running Software Version 3.4 & 3.4.1

Root Cause

Software design

Action Taken

Urgent Product Correction Notification letters dated 3/11/19 were sent to customers.