IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

This recall has been terminated (originally issued April 12, 2019).

Company
Topcon Medical Systems, Inc.
Recall Initiated
March 6, 2019
Posted
April 12, 2019
Terminated
July 17, 2019
Recall Number
Z-1138-2019
Quantity
450 units
Firm Location
Oakland, NJ
Official Source
View on FDA website ↗

Reason for Recall

When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.

Distribution

Nationwide

Lot / Code Info

System Version 1.53

Root Cause

Labeling design

Action Taken

Topcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at TMSservice@topcon.com