VIDAS LH, Reference Numbers 30406 and 30406-01. Poduct Usage: VIDAS LH is an automated quantitative test for use on the VIDAS family instruments for the determination of human luteinizing hormone in human serum or plasma (lithium heparin), using the ELFA technique (Enzyme Linked Fluorescent Assay)

Company
Biomerieux, Inc.
Recall Initiated
March 1, 2018
Terminated
December 9, 2025
Recall Number
Z-2588-2018
Quantity
5,429 units
Firm Location
Durham, NC

Reason for Recall

Several complaints were received for invalid calibration with low calibrator S1 while using VIDAS LH. Internal tests confirmed a decrease of the strips signal overtime for specific lots.

Distribution

Worldwide Distribution - US Nationwide in the state of Arizona and North Carolina. Distributed internationally to Algeria, Algeria, Angola, Argentina, Bahrain, Bangladesh, Belarus, Bosnia-Herz., Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chad, Colombia, Congo, Costa Rica, Cote d'Ivoire, Czech Republic, Djibouti, Dutch Antilles, Ecuador, Egypt, Estonia, Ethiopia, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Mali, Mauretania, Mexico, Morocco, Mozambique, Myanmar, Netherlands, Nicaragua, Nigeria, Pakistan, Palestine, Panama, Paraguay, Philippines, Poland, Portugal, Republic of Congo, Romania, Russia, Saudi Arabia, Senegal, Serbia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, Uruguay, United Arab Emirates.

Lot / Code Info

Reference No. 30406-01, Lot No. 1006041520 (Exp. Date 10/06/2018);   Reference No. 30406, Lot No. (Exp. Date): 1005940920 (08/06/2018), 1006041540 (10/06/2018), 1006212740 (12/21/2018), 1006274770 (01/29/2019)

Root Cause

Nonconforming Material/Component

Action Taken

bioMerieux notified customers on about 03/01/2018, via letter. Instructions included to distribute the recall information to all appropriate personnel in the laboratory, forward the information to individuals to which the product many have been transferred to, destroy any impacted lots on site, and complete and return the Acknowledgement Form. Questions and concerns can be directed to local bioMerieux Customer Service representatives or for Clinical Technical Support Center by email at CustomerService-ImmunoMolecular@biomerieux.com.