RecallDepth

Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label. Size (Part Number): a) D 2.0MM X 8MM (0002-D-20008); b) D 2.0MM X 10MM (002-D-20010); c) D 2.0MM X 12MM (002-D-20012); d) D 2.0MM X 14MM (002-D-20014); e) D 2.0MM X 16MM (002-D-20016); f) D 2.0MM X 18MM (002-D-20018); g) D 2.0MM X 20MM (002-D-20020); h) D 2.0MM X 22MM (002-D-20022); i) D 2.0MM X 24MM (002-D-20024); j) D 2.0MM X 26MM (002-D-20026); k) D 2.0MM X 28MM (002-D-20028); l) D 2.0MM X 30MM (002-D-20030)

This recall has been terminated (originally issued June 7, 2018).

Company
Orthofix, Inc
Recall Initiated
April 6, 2018
Posted
June 7, 2018
Terminated
June 30, 2021
Recall Number
Z-2119-2018
Quantity
601
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Lot / Code Info

(UDI)/Lot: a) (18056099645614) B1204542, B1101808;  b) (18056099645621) B1204969, B1201169, B1075101;  c) (18056099645638) B1204554, B1195701, B1075428;  d) (18056099645645) B1204971, B1201176, B1075846;  e) (18056099645652) B1195727, B1204266, B1075426;  f) (18056099645669) B1204540, B1075429;  g) (18056099645676) B1204541, B1075425;  h) (18056099645683) B1204539, B1075177;  i) (18056099645690) B1204972, B1195700, B1075847;  j) (18056099645706) B1204555, B1195702, B1075427;  k) (18056099645713) B1204973, B1201170, B1075430;  l) (18056099645720) B1075825, B1201166, B1204270

Root Cause

Device Design

Action Taken

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.

More recalls by Orthofix, Inc

View all recalls by this company →