Rival Reduce Kit Locking and Non-Locking Screw D 2.0MM, D 2.7MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-011C

This recall has been terminated (originally issued June 7, 2018).

Company
Orthofix, Inc
Recall Initiated
April 6, 2018
Posted
June 7, 2018
Terminated
June 30, 2021
Recall Number
Z-2116-2018
Quantity
50
Firm Location
Lewisville, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Distribution

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Lot / Code Info

UDI: 18056099648592,  B1115657,  B1107222,  B1107223

Root Cause

Device Design

Action Taken

On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.