Electrohydraulic lithotripter probe, 1.9 FR 375 cm, Model 9-195-371DS used with a Model 9-201-00 Autolith lithotripter. electrohydraulic lithotripsy (EHL) is a method that allows large stones to be fragmented and removed percutaneously or endoscopically.

Company
Northgate Technologies, Inc.
Recall Initiated
May 24, 2017
Terminated
June 9, 2020
Recall Number
Z-0073-2018
Quantity
186
Firm Location
Elgin, IL

Reason for Recall

The inner and outer labels on the product contain incorrect reference number of 9-202-3751DS. According to Northgate Technologies, the resistor in the probe will not allow the mislabeled product to be used with the incorrect generator. Could Delay Care.

Distribution

US Distribution to the following states : MA, NE and TN and Internationally to Canada.

Lot / Code Info

Model No. 9-195-371DS, Lot/Serial Number LDS12516

Root Cause

Labeling mix-ups

Action Taken

Northgate Technologies sent an Urgent Medical Device Recall Notification dated May 24, 2017. Instructions included to examine records and locate affected devices, cease distribution or use and quarantine any affected products, return any affected product to Northgate, and contact customers if the affected product has been further distributed. For further questions, please call (800) 348-0424.