RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile, Intermittent self-catheterization
This recall has been terminated (originally issued November 22, 2016).
- Company
- Teleflex Medical
- Recall Initiated
- November 22, 2016
- Terminated
- November 24, 2017
- Recall Number
- Z-0771-2017
- Quantity
- 102,843 units
- Firm Location
- Research Triangle Park, NC
- Official Source
- View on FDA website ↗
Reason for Recall
Labeling Inconsistency: The products have been labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.
Distribution
Nationwide Distribution to CA, CO, FL, GA, IL, KS, KY, MA, MO, MS, NC, NJ, NY, OH, OK, OR, PA and TX.
Lot / Code Info
Product Code: 10096080, Lot number: 74F1501853 (8 FR); Product Code: 10096120, Lot numbers: 74E1602795, 74F1503201, 74G1600723 & 74J1500576 (12 FR); Product Code: 10096140, Lot numbers: 74A1600023, 74A1601105, 74A1602234, 74D1602007, 74E1602235, 74E1602796, 74F1501855, 74G1600486, 74H1500530 & 74L1500363 (14 FR); Product Code: 10096100, Lot numbers: 74A1600799, 74A1600800, 74A1602236, 74A1603122, 74D1600231, 74D1600923, 74D1601540, 74D1602136, 74E1600845, 74E1601768, 74E1602237, 74F1501857, 74F1501962, 74F1601966, 74F1601988, 74H1500534, 74H1500978, 74J1501157, 74K1502306, 74L1500364, 74L1501303, 74L1501938, 74M1500227 and 74M1501087 (10 FR) and Product Code: 10096160, Lot numbers: 74A1600024, 4A1600801, 74A1601106, 74A1603123, 74D1601541, 74E1600846, 74E1602797, 74G1600724, 74H1500979, 74J1501158, 74K1501837, 74L1501202 &74L1501939 (16 FR).
Root Cause
Error in labeling
Action Taken
Teleflex sent an Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers with affected stock, immediately discontinue use and quarantine any affected products. To return product, customers were advised to complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recall@teleflex.com. Customers with no affected stock should also complete the enclosed Recall Acknowledgment Form to confirm receipt of the letter. Customers with questions should contact Customer Service at 1-866-246-6990.