RUSCH, Pocket PAC IC Intermittent Cath Kit, 8 FR, 12 FR, 14 FR, 10 FR and 16 FR, Rx Only, Sterile, Intermittent self-catheterization

This recall has been terminated (originally issued November 22, 2016).

Company
Teleflex Medical
Recall Initiated
November 22, 2016
Terminated
November 24, 2017
Recall Number
Z-0771-2017
Quantity
102,843 units
Firm Location
Research Triangle Park, NC
Official Source
View on FDA website ↗

Reason for Recall

Labeling Inconsistency: The products have been labeled with the incorrect expiration date which exceeds the actual, validated 35-month shelf life.

Distribution

Nationwide Distribution to CA, CO, FL, GA, IL, KS, KY, MA, MO, MS, NC, NJ, NY, OH, OK, OR, PA and TX.

Lot / Code Info

Product Code: 10096080, Lot number: 74F1501853 (8 FR); Product Code: 10096120, Lot numbers: 74E1602795, 74F1503201, 74G1600723 & 74J1500576 (12 FR); Product Code: 10096140, Lot numbers: 74A1600023, 74A1601105, 74A1602234, 74D1602007, 74E1602235, 74E1602796, 74F1501855, 74G1600486, 74H1500530 & 74L1500363 (14 FR); Product Code: 10096100, Lot numbers: 74A1600799, 74A1600800, 74A1602236, 74A1603122, 74D1600231, 74D1600923, 74D1601540, 74D1602136, 74E1600845, 74E1601768, 74E1602237, 74F1501857, 74F1501962, 74F1601966, 74F1601988, 74H1500534, 74H1500978, 74J1501157, 74K1502306, 74L1500364, 74L1501303, 74L1501938, 74M1500227 and 74M1501087 (10 FR) and Product Code: 10096160, Lot numbers: 74A1600024, 4A1600801, 74A1601106, 74A1603123, 74D1601541, 74E1600846, 74E1602797, 74G1600724, 74H1500979, 74J1501158, 74K1501837, 74L1501202 &74L1501939 (16 FR).

Root Cause

Error in labeling

Action Taken

Teleflex sent an Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers with affected stock, immediately discontinue use and quarantine any affected products. To return product, customers were advised to complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recall@teleflex.com. Customers with no affected stock should also complete the enclosed Recall Acknowledgment Form to confirm receipt of the letter. Customers with questions should contact Customer Service at 1-866-246-6990.