RecallDepth

.045 X 6 ST Guide Wire, Part number 45-0029-S, Sterile 10 pack of WS-1106ST and has a 6 inch length and 0.045 inch diameter. .054 X 7 ST Guide Wire, Part number 45-0031-S, Sterile 10 Pack of WS-1407ST and has a 7 inch length and 0.054 inch diameter. These guide wires are an orthopedic manual surgical instrument.

This recall has been terminated (originally issued March 16, 2017).

Company
Acumed Llc
Recall Initiated
February 17, 2017
Posted
March 16, 2017
Terminated
May 19, 2017
Recall Number
Z-1483-2017
Quantity
24 units of .045 X 6 ST Guide Wire and 25 units of .054 X 7 ST Guide Wire
Firm Location
Hillsboro, OR
Official Source
View on FDA website ↗

Reason for Recall

Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or nonunion, need for revision surgery, and histological response (if wire breaks).

Distribution

Worldwide Distribution in Great Britain, Ireland, and Japan only.

Lot / Code Info

.045 X 6 ST Guide Wire Lot number: 391681  .054 X 7 ST Guide Wire Lot number: 391680

Root Cause

Process control

Action Taken

Acumed sent an Urgent Medical Device Recall letter dated 02/16/2017, to all affected customers on 02/17/2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or call to obtain a Return Material Authorization (RMA) number. Customers with questions about recall can call 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada).