Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.

This recall has been terminated (originally issued October 31, 2016).

Company
Hill-Rom, Inc.
Recall Initiated
August 9, 2016
Posted
October 31, 2016
Terminated
March 24, 2017
Recall Number
Z-0326-2017
Quantity
188
Firm Location
Batesville, IN
Official Source
View on FDA website ↗

Reason for Recall

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

Distribution

WA NC OH PA MI Australia Sweden

Lot / Code Info

manufacturing date between Dec 01, 2015- Apr 14, 2016

Root Cause

Mixed-up of materials/components

Action Taken

On 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.