RecallDepth

MULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems The product is a medical device (chemistry reagent) and is intended for the quantitative determination of creatinine in human serum, plasma, and urine using the ARCHITECT c Systems. The kit is provided as a ready to use liquid material consisting in two components: Reagent 1 and Reagent 2. According to the catalogue number, the kit is composed of three vials of 48 ml for Reagent 1 and three vials of 18 ml for Reagent 2 in the configuration 8L24-31. The configuration 8L24-41 consists in ten vials of 90 ml for Reagent 1 and ten vials of 34 ml for Reagent 2

Company
Sentinel Ch Spa Via Robert Koch
Recall Initiated
May 22, 2017
Posted
June 27, 2017
Terminated
September 4, 2018
Recall Number
Z-2867-2017
Quantity
2356 kits Worldwide; US 206 kits

Reason for Recall

Reagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.

Distribution

Nationwide Distribution

Lot / Code Info

Catalog Number 8L24-31: Lot Numbers: 60335Y600, 60382Y600 (Exp. Date 2018-02-28); Catalog Number 8L24-41: Lot Number: 60334Y600, (Exp. Date 2018-02-28)

Root Cause

Nonconforming Material/Component

Action Taken

Sentinel Diagnostics notified the US Distributor on 5/22/17 to notify US customers by letter. Actions to be Taken 1. Discontinue the use of the impacted lots immediately and destroy any remaining inventory of impacted lots according to your laboratory procedures. 2. If you have an alternate reagent lot available in inventory, switch to this alternate lot, or use a new lot provided. 3. Please review the content of this communication with your Medical Director and retain this letter for any future reference. Questions regarding this information, U.S. Customers should contact Customer Service at 1-877-4 ABBOTT. Customers to complete Field Safety Notice Receipt.