CORFLO NG/NI Pediatric Neonatal Feeding Tube w/ Anti-IV Connector

This recall has been terminated (originally issued October 16, 2020).

Company
Avanos Medical, Inc.
Recall Initiated
October 16, 2020
Terminated
September 30, 2022
Recall Number
Z-0540-2021
Quantity
710 units
Firm Location
Alpharetta, GA
Official Source
View on FDA website ↗

Reason for Recall

Detachment of the Male AIV Adaptor from the Y-Connector.

Distribution

Domestic distribution only: CA, CO, IL, LA, OH, VA.

Lot / Code Info

Product Code:20-1225AIV, Lot Number: 0002983985

Root Cause

Under Investigation by firm

Action Taken

On 10/16/2020 the firm sent a notification to its consignees with the following instructions: - CHECK all storage and usage locations to determine if any impacted product remains within your possession. Product Currently In-Use -DISCONTINUE using the impacted product (Note: If you decide not to replace the device, please consider taping over the adaptor to prevent any potential detachment and be sure to check on the tube and adaptor frequently throughout the day for potential detachment). -REPLACE the impacted product with a unit of a different lot. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos . Via email to fieldactioncare@avanos.com -DESTROY the devices in your inventory following your facility's destruction procedures . Unused Inventory -SEGREGATE and QUARANTINE the devices according to your facility's procedures. -IMMEDIATELY COMPLETE and RETURN the attached Recall Acknowledgement Form to Avanos. Via email to fieldactioncare@avanos .com -DESTROY the devices in your inventory following your facility's destruction procedures.

More recalls by Avanos Medical, Inc.

View all recalls by this company →