Non-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm, REF W27 ST012 (c) DRP 0.27 - 12 mm, REF W27 ST012 (d) NEO(R) 0.27 - 14 mm, REF W27 ST014 (e) NEO(R) 0.27 - 16 mm, REF W27 ST016 (f) NEO(R) 0.27 - 18 mm, REF W27 ST018 (f) DRP 0.27 - 18 mm, REF W27 ST018 The NEOVIEW(R) Plating System is intended for fixation of intra-articular and extra-articular fractures of the distal radius and reconstruction of the distal radius. It is composed of a NEOVIEW" plate in PEEK and NE0411 screws in Ti6Al4V that are available in Locking or Non-Locking versions. The NEOVIEW411 Screw Set consists of a standard set of NE0411 screws with different lengths suitable for the implantation of a NEOVIEW9 plate for an average anatomy

This recall has been terminated (originally issued July 13, 2017).

Company
In2Bones, Sas 28 Chemin Du Petit Bois Ecully France
Recall Initiated
July 13, 2017
Terminated
August 27, 2018
Recall Number
Z-2869-2017
Quantity
287 screws
Official Source
View on FDA website ↗

Reason for Recall

The firm received several customer complaints reporting some perioperative breakages of NE0411 Non-Locking screws when inserting the screw into the radius diaphysis

Distribution

Worldwide Diribution - US including Tennessee; Foreign - Italy, Portugal, Spain, Switzerland, UK, and France

Lot / Code Info

All batch numbers

Root Cause

Device Design

Action Taken

In2Bones sent a Field Safety Corrective Action letter dated July 13, 2017, to all affected customers. The consignee was directed to remove the products from distribution and return it to the recalling firm. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at: +33 4 72 29 26 26/ +337 61 88 44 30 or by email: qualite@in2bones.com.