Freelite Human Lambda Free Kit for use on SPAPLUS Product Code: LK018.S, LK018.10S Freelite aids the detection and monitoring of Multiple Myeloma and related diseases

This recall has been terminated (originally issued January 19, 2017).

Company
The Binding Site Group, Ltd.
Recall Initiated
October 28, 2016
Posted
January 19, 2017
Terminated
July 27, 2017
Recall Number
Z-1053-2017
Quantity
3,088
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

A change in the calibration curve causing an increase in false prozone flags.

Distribution

Worldwide Distribution - U.S. (nationwide) and Internationally to the following; Canada, Australia.

Lot / Code Info

Lot #400012, 400257

Root Cause

Under Investigation by firm

Action Taken

Binding Site sent an Important Notice dated October 28, 2016, to all affected customers. . Binding Site requests the following: - Immediately quarantine and dispose of any LK018.S lot 400012 / LK018.10S lot 400257 kits held in your organisation; - Remove any affected lots from the SPAPLUS instrument; - Contact Binding Site representative to request replacement kits; - Return completed E-back form to your local Binding Site representative. For questions regarding this recall call 858-453-9177.