Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

This recall has been terminated (originally issued January 18, 2017).

Company
The Binding Site Group, Ltd.
Recall Initiated
July 5, 2016
Posted
January 18, 2017
Terminated
July 3, 2017
Recall Number
Z-1043-2017
Quantity
809 kits
Firm Location
Birmingham United Kingdom
Official Source
View on FDA website ↗

Reason for Recall

Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Distribution

U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.

Lot / Code Info

Lot #405566, 405567

Root Cause

Under Investigation by firm

Action Taken

On 07/05/16 notification letters were sent to customers. The Binding Site Group requests the following: - Immediately Quarantine & Dispose of any LK018.OPT lot 405566 kits within stock; - Remove any affected lots from the Optilite instrument; - Patients with Lambda Freelite results >114mg/L should be reviewed in consultation with clinicians, alongside other clinical findings and retested where applicable; - Notify local Binding Site representative to request replacement kits; - Return completed E-Back Form to local Binding Site Representative; The firm recommended that the notice be circulated to those within the organization who need to be aware.